INSTRUCTION MANUAL CS300 Pyranometer Revision: 1/15 Copyright © 1994- 2015 Campbell Scientific, Inc.
CS300 Pyranometer 2. The Short Cut installation should place a Short Cut icon on the desktop of your computer. To open Short Cut, click on this icon
CS300 Pyranometer 5. Under the Available Sensors and Devices list, select Sensors | Meteorological | Solar Radiation folder. Select CS300 Pyranomete
CS300 Pyranometer 7. Select any other sensors you have, then finish the remaining Short Cut steps to complete the program. The remaining steps are o
CS300 Pyranometer Cosine response: ±5% at 75° zenith angle. ±2% at 45° zenith angle Response time: < 1 ms Temperature response: < 1% at 5
CS300 Pyranometer 7.2 Mounting to an Instrument Mount 7.2.1 Required Tools Tools required for installation on a tripod or tower: Small and medium Ph
CS300 Pyranometer 3. Place the CS300 in the center of the 18356 base/leveling fixture. 4. Loosely mount the 18356 base/leveling fixture on the CM2
CS300 Pyranometer 7. Route the sensor cable along the underside of the crossarm to the tripod/tower, and to the instrument enclosure. 8. Secure the
CS300 Pyranometer 3. Loosely mount the 18356 base/leveling fixture on the 015ARM. Do not fully tighten the three mounting screws. 4. Turn the
CS300 Pyranometer TABLE 7-1. Connections to Campbell Scientific Dataloggers Color Wire Label CR6 CR9000(X) CR5000 CR3000 CR1000 CR800 CR8
CS300 Pyranometer TABLE 7-2. Multipliers Required for Average Flux and Total Flux Density in SI and English Units UNITS MULTIPLIER PROCESS W m-2 5
CS300 Pyranometer 8. Maintenance and Calibration On a monthly basis the level of the pyranometer should be checked. Any dust or debris on the sensor
Appendix A. Importing Short Cut Code Into CRBasic or Edlog Editor This tutorial shows: • How to import a Short Cut program into a program editor fo
Appendix A. Importing Short Cut Code Into CRBasic or Edlog Editor Once the file is edited with CRBasic Editor, Short Cut can no longer be used to
Appendix B. Example Programs The following programs measure the CS300 every 10 s and convert the millivolt output to Wm–2 and MJm–2. Both programs
Appendix B. Example Programs B.2 CR10X Example Program ;{CR10X} *Table 1 Program 01: 10.0000 Execution Interval (seconds) ; Measure CS300 pyr
Appendix B. Example Programs 12: Set Active Storage Area (P80) 1: 1 Final Storage Area 1 2: 102 Array ID 13: Real Time (P77)
Appendix B. Example Programs B-4
Campbell Scientific Companies Campbell Scientific, Inc. (CSI) 815 West 1800 North Logan, Utah 84321 UNITED STATES www.campbellsci.com • info@campbe
Limited Warranty “Products manufactured by CSI are warranted by CSI to be free from defects in materials and workmanship under normal use and service
Assistance Products may not be returned without prior authorization. The following contact information is for US and international customers residing
Precautions DANGER — MANY HAZARDS ARE ASSOCIATED WITH INSTALLING, USING, MAINTAINING, AND WORKING ON OR AROUND TRIPODS, TOWERS, AND ANY ATTACHMENTS TO
Table of Contents PDF viewers: These page numbers refer to the printed version of this document. Use the PDF reader bookmarks tab for links to spec
Table of Contents Tables 7-1. Connections to Campbell Scientific Dataloggers ... 10 7-2. Multipliers Required for Average
CS300 Pyranometer 1. Introduction The CS300, manufactured by Apogee Instruments, measures total sun and sky solar radiation for solar, agricultural,