AM16/32 Relay Analog Multiplexer
17
*1 Table 1 Programs
1: 60 Sec. Execution Interval
ACTIVATES MULTIPLEXER
1: Do (P86)
1: 41 Set high Port 1
BEGINS STRAIN GAGE MEASUREMENT LOOP
2: Beginning of Loop (P87)
1: 0 Delay
2: 10 Loop Count
CLOCK PULSE
3: Do (P86)
1: 72 Pulse Port 2
DELAY
4: Excitation with Delay (P22)
1: 1 EX Chan
2: 0 Delay w/EX (units=.01sec)
3: 1 Delay after EX (units=.01sec)
4: 0 mV Excitation
FULL BRIDGE MEASUREMENT INSTRUCTION
5: Full Bridge (P6)
1: 1 Rep
2: 3 50 mV slow Range
3: 1 IN Chan
4: 1 Excite all reps w/Enchain 1
5: 5000 mV Excitation
6: 1-- Loc [:STRAIN #1]
7: 1 Mult
8: 0 Offset
END OF STRAIN GAGE MEASUREMENT LOOP
6: End (P95)
BEGINNING OF POTENTIOMETER MEASUREMENT LOOP
7: Beginning of Loop (P87)
1: 0 Delay
2: 6 Loop Count
8: Step Loop Index (Extended) (P90)
1: 2 Step
CLOCK PULSE
9: Do (P86)
1: 72 Pulse Port 2