CNR4 Net Radiometer
7.2.2 Example 2, CR3000 Program Using Differential Measurements
The Program Example 2 requires four differential channels to measure the four
radiation outputs and one excitation channel and one single-ended channel to
measure the thermistor. The program measures the sensors every 1 second,
performs the on-line processing of the data and stores the following processed
data to a data table called cnr4_data once every 60 minutes. It also stores the
raw time-series data from CNR4 to data table called cnr4_ts.
Minimum Battery voltage
Sample Datalogger panel temperature
Average Short-wave radiation (pyranometer up)
Average Short-wave radiation (pyranometer down)
Average Long-wave radiation (pyrgeometer up)
Average Long-wave radiation (pyrgeometer down)
Average CNR4 thermistor temperature (degrees C)
Average CNR4 thermistor temperature (Kelvin)
Average Corrected long-wave radiation (pyrgeometer up)
Average Corrected long-wave radiation (pyrgeometer down)
Average Short-wave net radiation
Average Long-wave net radiation
Average Albedo
Average Net radiation
'CR3000 Series Datalogger
'
'CNR4 program
'This program measures CNR4 four-component net radiometer
'This program also measures the thermistor inside the CNR4
'
'User must enter the sensitivity values for all four probes in the program and save/compile
'prior to downloading it to the datalogger.
'Search for the text string "unique" to find places to enter the sensitivity values.
'
'Wiring Instructions
'
'ANALOG CHANNELS
'1H CNR4 Pyranometer Upper signal (red)
'1L CNR4 Pyranometer Upper signal reference (blue)
'gnd jumper to 1L
'
'2H CNR4 Pyranometer Lower signal (white)
'2L CNR4 Pyranometer Lower signal reference (black)
'gnd jumnper to 2L
'
'3H CNR4 Pyrgeometer Upper signal (grey)
'3L CNR4 Pyrgeometer Upper signal reference (yellow)
'gnd jumper to 3L
'
'4H CNR4 Pyrgeometer Lower signal (brown)
'4L CNR4 Pyrgeometer Lower signal reference (green)
'gnd jumper to 4L
' CNR4 shield (clear)
'
'
21