TD ADDENDUM—SECTION 8. PROCESSING AND PROGRAM CONTROL EXAMPLES
AD-8-8
Time into Test, min Output Interval Loop #
00 to 10 10 sec. 1
10 to 30 30 sec. 2
30 to 100 1 min. 3
100 to 300 2 min. 4
300 to 1000 5 min. 5
1000 and greater 10 min. 6
This is accomplished with a series of loops
(
Instruction 87), where the delay and count
parameters are used to implement the
frequency of measurement (and output) and the
duration of the that frequency. The unit of delay
is the execution interval. A delay of 1 with a 10
second execution interval and a count of 60
means the instructions in the loop, in this case
measure and output water level, are executed
every 10 seconds for 10 minutes.
The drawdown portion of the test is completed
at s
ome time greater than 1000 minutes. To
enter the recharge phase of the test, the
operator enters the *6AD Mode and sets Flag 1
high. At the next 10 minute pass through loop 6
the loop is exited. Program execution returns to
the top of the program table and the
measurement schedule starts over again for the
recharge test.
The sensor is a 50 PSI Druck, model 930/ti with
a c
alibration of 49.93 mV/10V of excitation or
4.993mV/V. Your calibration will be different.
An excitation voltage of 1500 mV yields a
maximum signal of 7.489 mV at 50 PSI, fully
utilizing the 7.5 mV Input Range to provide the
best resolution.
The multiplier, m, is calculated to provide depth
of
water in feet:
m = (50 psi/4.993 mV/V) * (2.3067 ft/psi)
m = 23.099 ft/mV/V
The offset is calculated to provide a final value
that r
epresents the distance from the lip of the
well to the water surface. Similar to Figure
7.16-2, the offset equals the initial distance of
47.23 feet plus the initial reading of 54.77, or
102 feet.
* 1 Table 1 Programs
01: 10 Sec. Execution Interval
User must toggle Flag 1 to start measurements
01: P91 If Flag/Port
01: 21 Do if flag 1 is low
02: 0 Go to end of Program Table
Loop 1, Output every 10 seconds for 10
minutes
02: P87 Beginning of Loop
01: 1 Delay
02: 60 Loop Count
03: P86 Do
01: 1 Call Subroutine 1
04: P95 End
Loop2, Output every 30 seconds for 20 minutes
05: P87 Beginning of Loop
01: 3 Delay
02: 40 Loop Count
06: P86 Do
01: 1 Call Subroutine 1
07: P95 End
Loop 3, Output every 1 minute for 70 minutes
08: P87 Beginning of Loop
01: 6 Delay
02: 70 Loop Count
09: P86 Do
01: 1 Call Subroutine 1
10: P95 End
Loop 4, Output every 2 minutes for 200 minutes
11: P87 Beginning of Loop
01: 12 Delay
02: 100 Loop Count
12: P86 Do
01: 1 Call Subroutine 1
13: P95 End
Loop 5, Output every 5 minutes for 700 minutes
14: P87 Beginning of Loop
01: 30 Delay
02: 140 Loop Count
15: P86 Do
01: 1 Call Subroutine 1
16: P95 End