Campbell CS451 Manuel Page 1

Naviguer en ligne ou télécharger Manuel pour Mesure, test et contrôle Campbell CS451. Campbell CS451 Product manual Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 36
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 0
CS451/CS456 Submersible
Pressure Transducer
Revision: 1/13
Copyright © 2009-2013
Campbell Scientific, Inc.
Vue de la page 0
1 2 3 4 5 6 ... 35 36

Résumé du contenu

Page 1 - Pressure Transducer

CS451/CS456 Submersible Pressure TransducerRevision: 1/13Copyright © 2009-2013Campbell Scientific, Inc.

Page 2

CS451/CS456 Submersible Pressure Transducer 4.6 Step 6 — Use SCWin to Program Datalogger and Generate Wiring Diagram The simplest method for programm

Page 3 - Warranty

CS451/CS456 Submersible Pressure Transducer 3. Select CS451/CS456 Pressure Sensor, click the right arrow to add it to the list of sensors to be measu

Page 4 - Assistance

CS451/CS456 Submersible Pressure Transducer 8. Click on Wiring Diagram and wire according to the wiring diagram generated by SCWin Short Cut. Refer

Page 5 - Table of Contents

CS451/CS456 Submersible Pressure Transducer The CS451/CS456 has two communication options: SDI-12 or RS-232. The CS451/CS456 is shipped from the fact

Page 6

CS451/CS456 Submersible Pressure Transducer 6. Specifications Features: • Output acceptable for recording devices with SDI-12 or RS-232 capability

Page 7 - Transducer

CS451/CS456 Submersible Pressure Transducer Overpressure: 2x pressure range Dry Storage Temperature: ─10° to 80°C Sensor will be damaged if it is e

Page 8 - 4. Quickstart

CS451/CS456 Submersible Pressure Transducer 7. Operation 7.1 Configuration The CS451/CS456 is configured at the factory with default settings: TABLE

Page 9

CS451/CS456 Submersible Pressure Transducer 7.1.1.3 Powering the Sensor The A200 provides power to the sensor when it is connected to a PC’s USB port

Page 10 - Wiring Diagram

CS451/CS456 Submersible Pressure Transducer 7.1.2 Device Configuration Utility (version 2.03 or higher) The Device Configuration Utility allows you t

Page 11

CS451/CS456 Submersible Pressure Transducer Transparent mode is entered while the PC is in telecommunications with the datalogger through a terminal e

Page 13

CS451/CS456 Submersible Pressure Transducer Now check for response from the sensor with address zero by typing the SDI-12 Identify command 0I!<Ente

Page 14 - 6. Specifications

CS451/CS456 Submersible Pressure Transducer FIGURE 7-4. Terminal Emulator TABLE 7-4. RS-232 Terminal Commands Terminal Commands Values Returned 1

Page 15

CS451/CS456 Submersible Pressure Transducer TABLE 7-5. SDI-12 Commands SDI-12 Command Command Function Values Returned aM! Configured settings Pre

Page 16 - 7. Operation

CS451/CS456 Submersible Pressure Transducer address. If additional control ports are not available, then the address will need to be changed. 7.1.5.

Page 17

CS451/CS456 Submersible Pressure Transducer 7.2 Installation Considerations The CS451/CS456 is designed for water level measurements. Typical applic

Page 18

CS451/CS456 Submersible Pressure Transducer 7.2.3 Split Mesh Cable Grip (pn 25431) The 25431 Split Mesh Cable Grip is recommended for use in wells to

Page 19

CS451/CS456 Submersible Pressure Transducer 7.3.1 SDI-12 TABLE 7-6. SDI-12 Wiring Color CS451/CS456Function CR800 CR5000 CR3000 CR1000 CR200(

Page 20

CS451/CS456 Submersible Pressure Transducer 7.4.1 CRBasic Programming Dataloggers that use CRBasic include our CR200(X)-series, CR800, CR850, CR1000,

Page 21 - 7.1.5 SDI-12 Commands

CS451/CS456 Submersible Pressure Transducer 7.4.1.2 Example Program for CR1000 Datalogger 'CR1000 Series Datalogger 'Declare the variable

Page 22

CS451/CS456 Submersible Pressure Transducer 7.4.2.1 Example Program for CR10(X) Dataloggers Below is a portion of a CR10X program that measures the C

Page 23

Warranty “PRODUCTS MANUFACTURED BY CAMPBELL SCIENTIFIC, INC. are warranted by Campbell Scientific, Inc. (“Campbell”) to be free from defects in materi

Page 24 - 7.2.2 Offset Calculation

CS451/CS456 Submersible Pressure Transducer • The desiccant in the tube changes color from blue to pink when the drying power is lost. • The Enclosu

Page 25 - 7.3 Wiring

CS451/CS456 Submersible Pressure Transducer Suggestion: A kinked or plugged vent tube will not effectively vent a gauge pressure (Vented) type of devi

Page 26 - 7.4 Programming

CS451/CS456 Submersible Pressure Transducer 26

Page 27 - 7.4.1 CRBasic Programming

Appendix A. Calibration Certificate Each CS451/CS456 has been calibrated to meet printed accuracy specification at multiple temperature and pressure

Page 28 - 7.4.2 Edlog Programming

Appendix A. Calibration Certificate A-2

Page 30 - 9. Troubleshooting

Campbell Scientific Companies Campbell Scientific, Inc. (CSI) 815 West 1800 North Logan, Utah 84321 UNITED STATES www.campbellsci.com • info@campbel

Page 31

Assistance Products may not be returned without prior authorization. The following contact information is for US and international customers residing

Page 32

Table of Contents PDF viewers: These page numbers refer to the printed version of this document. Use the PDF reader bookmarks tab for links to spec

Page 33

Table of Contents ii 7.4 Programming... 20 7.4.1 CRBasic Programmi

Page 34

CS451/CS456 Submersible Pressure Transducer 1. Introduction The CS451/CS456 Submersible Pressure Transducer provides pressure and temperature measur

Page 35

CS451/CS456 Submersible Pressure Transducer • The model number and pressure range is etched on the housing and the cable length is printed on the lab

Page 36

CS451/CS456 Submersible Pressure Transducer 4.3 Step 3 — Dislodge Bubbles While submersing the transducer, air bubbles may become trapped between the

Commentaires sur ces manuels

Pas de commentaire