Campbell CS455 Manuel Page 26

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 34
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 25
CS450/CS455 Pressure Transducer
7.1 Every Visit
Collect data
Visually inspect wiring and physical conditions
Check indicating desiccant or enclosure humidity indicator; service if
necessary
Check battery condition (inspect physical appearance and use a keyboard
display, PDA, or laptop to view the battery voltage)
Check all sensor readings: adjust transducer offsets if necessary
Check recent data
7.2 Every Two to Three Years or on a Rotating Schedule
Send the CS450/CS455 in for inspection and have them served and
or/replaced as needed.
8. Troubleshooting
The most common causes for erroneous pressure transducer data include:
poor sensor connections to the datalogger
damaged cables
damaged transducers
moisture in the vent tube
Problem:
Unit will not respond when attempting serial communications.
Suggestion:
Check the power (Red is +V and Black is Ground) and signal (White is SDI-12
Data) lines to ensure proper connection to the logger. Check the logger
program to ensure that the same port the SDI-12 data line is connected to is
specified in the measurement instruction.
Problem:
Transducer appears to be operating properly but data shows a periodic or
cyclic fluctuation not attributable to water level changes.
20
Vue de la page 25
1 2 ... 21 22 23 24 25 26 27 28 29 30 31 32 33 34

Commentaires sur ces manuels

Pas de commentaire