Campbell FW05 Manuel

Naviguer en ligne ou télécharger Manuel pour Mesure, test et contrôle Campbell FW05. Campbell FW05 Product manual Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer

Résumé du contenu

Page 1 - Sonic Anemometer

CSAT3 Three DimensionalSonic AnemometerRevision: 9/12Copyright © 1998-2012Campbell Scientific, Inc.

Page 2

CSAT3 Three Dimensional Sonic Anemometer 2 to apply the speed of sound correction described by Liu et al., 2001. The CSAT3 can be configured to make

Page 3 - Warranty

CSAT3 Three Dimensional Sonic Anemometer 3 REPORTING RANGE: SDM and RS-232 digital Outputs: Full scale wind: ±65.535 m s-1 autoranging between four r

Page 4 - Assistance

CSAT3 Three Dimensional Sonic Anemometer 4 2.4 Power Requirements VOLTAGE SUPPLY: 10 to 16 Vdc POWER: 2.4 W @ 60 Hz measurement frequency 1.2 W @ 2

Page 5 - Table of Contents

CSAT3 Three Dimensional Sonic Anemometer 5 3.2 DM Address Each CSAT3 on the SDM bus must have a unique address. A hex thumb switch within the electr

Page 6

CSAT3 Three Dimensional Sonic Anemometer 6 4.2 Mounting The current mounting configuration for the CSAT3 includes a CM250 Leveling Mount, which shoul

Page 7

CSAT3 Three Dimensional Sonic Anemometer 7 10.00 cmYZXLogan, Utah CSAT33-D SONIC ANEMOMETERS/N: 102MADE IN USA60o FIGURE 4-2. CSAT3 coordinate system

Page 8

CSAT3 Three Dimensional Sonic Anemometer 8 10.00 cmYZXLogan, Utah CSAT33-D SONIC ANEMOMETERS/N: 102MADE IN USA60o FIGURE 4-4. CSAT3 coordinate system

Page 9 - Anemometer

CSAT3 Three Dimensional Sonic Anemometer 9 The CSAT3 electronics box contains unique calibration information for its respective sensor head. Do not m

Page 10 - 2.2 Output Signals

CSAT3 Three Dimensional Sonic Anemometer 10 TABLE 5-4. CSAT3 SDM Output to a 21X Datalogger Channel Description Color Pin 1H SDM Data Green D C1

Page 11 - 2.3 Physical Description

CSAT3 Three Dimensional Sonic Anemometer 11 TABLE 5-8. FW05/FWC-L35 Fine Wire Thermocouple Description Color Signal High Purple Signal Low Red 6

Page 13 - 4. Installation

CSAT3 Three Dimensional Sonic Anemometer 12 TABLE 6-2. SDM Compatible Dataloggers Datalogger Model Datalogger PROMs or Operating System (OS) Require

Page 14 - 4.2 Mounting

CSAT3 Three Dimensional Sonic Anemometer 13 6.2 RS-232 Output The Windows PC support software (CSAT32.EXE) can be used to graph the CSAT3 data in rea

Page 15 - 10.00 cm

CSAT3 Three Dimensional Sonic Anemometer 14 signal processing techniques result in more accurate, lower noise measurements. Unlike CSAT3s running embe

Page 16 - 5. Wiring

CSAT3 Three Dimensional Sonic Anemometer 15 only by the spacing between transducers but also by the current wind speed and speed of sound conditions.

Page 17

CSAT3 Three Dimensional Sonic Anemometer 16 8. CSAT3 Trigger The CSAT3 takes a measurement of wind speed and the speed of sound when triggered from o

Page 18

CSAT3 Three Dimensional Sonic Anemometer 17 FIGURE 8-1. Typical average CSAT3 current drain at +12 VDC MiMi + 1Mi + 2Mi + 3Pi - 1Si - 2PiSi - 1Pi +

Page 19 - 6. CSAT3 Outputs

CSAT3 Three Dimensional Sonic Anemometer 18 8.1.1 Single-Measurement Mode In the single-measurement mode, the anemometer makes one measurement per tr

Page 20 - 6.1.3 CR9000

CSAT3 Three Dimensional Sonic Anemometer 19 10. Datalogger Programming using SDM The CRBasic CSAT3 () and EDLOG SDM-CSAT3 (P107) instructions are Inp

Page 21 - 7. Operational Principles

CSAT3 Three Dimensional Sonic Anemometer 20 SDMAddress/SDM Address: The unique SDM Address of the CSAT3 connected to the datalogger. In CRBasic data

Page 22 - 7.2 Embedded Code Version 3

CSAT3 Three Dimensional Sonic Anemometer 21 Command/Option 90, 91, 98, or 99 must not be used to collect data from a CSAT3 running a cold shifted cal

Page 23

Warranty “PRODUCTS MANUFACTURED BY CAMPBELL SCIENTIFIC, INC. are warranted by Campbell Scientific, Inc. (“Campbell”) to be free from defects in materi

Page 24 - 8. CSAT3 Trigger

CSAT3 Three Dimensional Sonic Anemometer 22 Ux Input Location: The Input Location to store the first of five data values to an EDLOG datalogger. The

Page 25 - 8.1 Sampling Modes

CSAT3 Three Dimensional Sonic Anemometer 23 data condition also occurs when the datalogger’s execution frequency (inverse of the execution interval) i

Page 26 - 9. Data Pipeline Delay

CSAT3 Three Dimensional Sonic Anemometer 24 TABLE 10-7. Wiring for Program Example 1 Description Color CR3000 SDM Data Green SDM-C1 SDM Clock Wh

Page 27

CSAT3 Three Dimensional Sonic Anemometer 25 ' ' ' (N) ' {-y} '

Page 28

CSAT3 Three Dimensional Sonic Anemometer 26 'Wind directions and speed. Dim wnd_out(8) 'Alias wnd_out(1) = wnd_spd - in compass coordinate s

Page 29

CSAT3 Three Dimensional Sonic Anemometer 27 Totalize (1,n,IEEE4,NOT (del_T_f) OR NOT (disable_flag_on(2))) FieldNames ("del_T_f_Tot") Tot

Page 30 - 10.4 Diagnostic Word

CSAT3 Three Dimensional Sonic Anemometer 28 'Break up the four CSAT3 warning flags into four separate bits. diag_csat_work = diag_csat de

Page 31

CSAT3 Three Dimensional Sonic Anemometer 29 TABLE 10-8. Wiring for Program Example 2 Description Color CR3000 SDM Data Green SDM-C1 SDM Clock W

Page 32

CSAT3 Three Dimensional Sonic Anemometer 30 ;If the CSAT3 head is pointing West of North (above figure), ;theta is negative. If the CSAT3 head is poin

Page 33

CSAT3 Three Dimensional Sonic Anemometer 31 ;Disable Intermediate Processing if any of the diagnostic flags are ;on,vi.e. diag >= 4096 or any of th

Page 34

Assistance Products may not be returned without prior authorization. The following contact information is for US and international customers residing

Page 35

CSAT3 Three Dimensional Sonic Anemometer 32 17: Real Time (P77) 1: 110 Day,Hour/Minute (midnight = 0000) 18: Resolution (P78) 1: 1 High Res

Page 36

CSAT3 Three Dimensional Sonic Anemometer 33 ;Load the angle between North and the CSAT3 head. ;The station operator must enter the angle theta. ; 4:

Page 37

CSAT3 Three Dimensional Sonic Anemometer 34 ;Configure CSAT3 with 10 Hz parameters. ; 12: SDM-CSAT3 (P107) 1: 1 Reps 2: 3 SDM Address 3: 10

Page 38

CSAT3 Three Dimensional Sonic Anemometer 35 24: Z=X (P31) 1: 5 X Loc [ diag ] 2: 30 Z Loc [ diag_work ] 25: If (X<=>F) (P89)

Page 39

CSAT3 Three Dimensional Sonic Anemometer 36 36: End (P95) 37: End (P95) ;Process CSAT3 Output data. ; 38: Beginning of Subroutine (P85) 1: 3

Page 40

CSAT3 Three Dimensional Sonic Anemometer 37 46: End (P95) 47: End (P95) End Program -Input Locations- 1 Ux 2 Uy 3 Uz 4 Ts 5 diag 6 _________ 7 __

Page 41

CSAT3 Three Dimensional Sonic Anemometer 38 Lightly dab the face of the transducers to remove water droplets. Applying excessive force on the face of

Page 42

CSAT3 Three Dimensional Sonic Anemometer 39 INSTANT ADHESIVE FIGURE 11-2. P/N 28902 CSAT3 Sonic Wick Spares Kit contents 11.2 Calibration The CSAT3

Page 43

CSAT3 Three Dimensional Sonic Anemometer 40 The standard datalogger instruction reports either the speed of sound (c), option 90 or 99, or the sonic t

Page 44

CSAT3 Three Dimensional Sonic Anemometer 41 FIGURE 11-4. Recommended settings for the wind vertical axis FIGURE 11-5. Recommended settings for the

Page 45 - 11. Maintenance

i Table of Contents PDF viewers: These page numbers refer to the printed version of this document. Use the PDF reader bookmarks tab for links to sp

Page 46 - 11.1 Wicks

CSAT3 Three Dimensional Sonic Anemometer 42 FIGURE 11-6. Wind data from CSAT3 s/n 315; wind offsets within specifications at 19°C 12. Select Backdo

Page 47 - 11.2 Calibration

CSAT3 Three Dimensional Sonic Anemometer 43 After receiving the “??” command, the CSAT3 will respond with text similar to that below for CSAT3s runnin

Page 48 - 11.2.2 Test for Wind Offset

CSAT3 Three Dimensional Sonic Anemometer 44 be changed if RI=1 (always power the RS-232 drivers). Check the setting of RI in the Long Status Message.

Page 49 - be less than ± 4 cm

A-1 Appendix A. CSAT3 Orientation A.1 Determining True North and Sensor Orientation The orientation of the CSAT3 negative x-axis is found by readin

Page 50

Appendix A. CSAT3 Orientation A-2 Declination angles are always subtracted from the compass reading to find True North. A declination angle East of T

Page 51

Appendix A. CSAT3 Orientation A-3 A.2 Online Magnetic Declination Calculator The magnetic declination calculator web calculator published by NOAA’s

Page 52

Appendix A. CSAT3 Orientation A-4

Page 53

B-1 Appendix B. Serial Communications with the CSAT3 B.1 CSAT3 RS-232 Commands The CSAT3 powers up its RS-232 drivers when it detects that the RTS

Page 54

Appendix B. Serial Communications with the CSAT3 B-2 TABLE B-1. RS-232 Command Codes Command Trigger before Command Description Trigger afterComm

Page 55

Appendix B. Serial Communications with the CSAT3 B-3 Nassi-Shneiderman diagrams are given in FIGURE B-1and FIGURE B-2 for collecting data using a PC

Page 56

Table of Contents ii 10. Datalogger Programming using SDM ...19 10.1 CRBasic CSAT3 () Instruction...

Page 57 - B.1 CSAT3 RS-232 Commands

Appendix B. Serial Communications with the CSAT3 B-4 TABLE B-2. RS-232 Output Word 0 Word 1 Word 2 Word 3 Word 4 b15 - b0 b15 - b0 b15 - b0

Page 58

Appendix B. Serial Communications with the CSAT3 B-5 TABLE B-6. Converting Speed of Sound from Word 3 Conversion Output Resolution (mm s-1) (°C @

Page 59

Appendix B. Serial Communications with the CSAT3 B-6 TABLE B-9. RS-232 Status, Response to the S or P Command Byte Description Value 0 Trigger S

Page 60 - B.3 Words 0 through 3

C-1 Appendix C. CSAT3 Measurement Theory C.1 Theory of Operation C.1.1 Wind Speed Each axis of the CSAT3 pulses two ultrasonic signals in opposite

Page 61 - B.4 Word 4

Appendix C. CSAT3 Measurement Theory C-2 blowing perpendicular to the sonic path. No additional off-line corrections are required as suggested by Li

Page 62

Appendix C. CSAT3 Measurement Theory C-3 REFERENCES Kaimal, J. C. and Businger, J. A.: 1963, “A Continuous Wave Sonic Anemometer-Thermometer”, J. Ap

Page 63 - C.1 Theory of Operation

Appendix C. CSAT3 Measurement Theory C-4

Page 64 - 1051== +γγ(.)

D-1 Appendix D. SDM Communications and Long Signal Cables D.1 Brief Description of SDM Clock Rates The fastest rate that a datalogger and SDM senso

Page 65

Appendix D. SDM Communications and Long Signal Cables D-2 D.2 CR3000 Example Connect all the SDM devices to the datalogger with the cable in questio

Page 66

Appendix D. SDM Communications and Long Signal Cables D-3 Take the following case, a 575 foot SDM signal and power cable is used to connect a CR23X d

Page 67 - Long Signal Cables

Table of Contents iii 4-4. CSAT3 coordinate system and original noncaptive mounting hardware (s/n 0107 to 0630)...

Page 68 - D.3 CR23X Example

Appendix D. SDM Communications and Long Signal Cables D-4

Page 70

Campbell Scientific Companies Campbell Scientific, Inc. (CSI) 815 West 1800 North Logan, Utah 84321 UNITED STATES www.campbellsci.com • info@campbel

Page 71

Table of Contents iv D-1. Data from Example CR3000 SDM Clock Rate Search Program... D-2 D-2. Data from Example CR23X SDM Clock Rate Search Program

Page 72

1 CSAT3 Three Dimensional Sonic Anemometer 1. General The CSAT3 is an ultrasonic anemometer for measuring wind speed in three dimensions. It uses t

Commentaires sur ces manuels

Pas de commentaire