HMP45C Temperature and RH Probe
4
5. Continue to carefully insert the probe a further 10mm, until the shoulder of
the larger section is approximately level with the hole in the lowest curved
disc, as shown in Figure 2 (b), below.
6. Tighten the large retaining nut by hand (do not use any tools) until the probe
is firmly gripped. If the probe does not appear to be sitting squarely inside
the shield, loosen the nut, re-position the probe and re-tighten. DO NOT
PULL THE PROBE DOWNWARDS PRIOR TO TIGHTENING THE
NUT, as this will also pull the sealing membrane downwards producing a
water trap. (See CAUTION below.)
7. Finally, bend the cable in a gentle loop back up towards the mounting arm
and attach it to the arm with the cable ties supplied.
Figure 2 Inserting the Probe into the Radiation Shield
Larger section of probe body in final
position (level with lip of first curved disc)
and showing the sealing membrane on
the flat disc in the correct
p
osition.
Sealing membrane must curve
upwards to repel water.
Flat white disc.
Carefully push the HMP45C up into
the radiation shield.
(a)
(b)
Observe here