
FIGURE
tg.4-2. Diagram of Junction Box
An external
reference
junction
box
must
be
constructed
so that the entire terminal area
is
very close tO the
same
temperature.
This
is
necessary sO that a valid
reference temperature
can
be measured. and to avoid
a thermoelectric
otfset
voltagB
which
will be
induced
if
the
terminals
at
Which
the
thermocouple
leads
are
connected
(points
A and
B in Figure
13.4-3)
are
at different
tpmperatures. The box should
contain
elernents
of
high thermal
conductivity,
which will adt to
rapidly remove any thermal
gradients
to
which
the box
is subjected. lt is not
necessary tQ design a constant
temperature
box;
it is de$irable
that
the box
respond
slowly
to external
temperatu re f luctuations.
Radiation
shielding
m{st be
provided
when a
junction
box
is installed
ifi the
field.
Care must
also
be taken
that
a thermpl
gradient
is
not induced by
conduction
frrough
the
incoming wires. The
21X
can
be
Used
to measure the
temperature
gradients
wifhin the
junction
box.
13.5 BRIDGE
RESISTANCE
MEASUREMENTS
There are 6
bridge measurement
instructions
included
in
the
standard 21X software. Figure
13.5-1 shovr4s
the circuits
that
would typically be
measured with
these instructions.
In
the
diagrams,
t(e
resistors labeled R"
would
normally be the sensors and those
labeled R1
would
normhlly be
fixed
resistors. Circuits other
than those
{iagrammed
could be
measured,
provided
th{
excitation
and
type of
measu
remefrts
were
appropriate.
With the exOeption of
Instructions
4
and 8,
which apply an
excitation voltage then wait a
specified
tirne before making
a measurement,
SECTION
13. 21X MEASUREIT,IENTS
allof the bridge
measurements make
one
set of
measurements with the
excitation
as
programmed
and
another
sel
of measurements
with
the
excitation
polarity
reversed.
The
error
in
the two
measurements due to
thermal emfs
is
then accounted
for in
the
processing
of
the
measurement instruction. The excitation
is
switched on
450ps before
the
integration
portion
of
the
measuremenl
stads
and
is
grounded
as
soon
as the
integration is
completed.
Figure
13.5-2
shows
the excitation and
measurement
sequence for
Instruction
6, a
4 wire full bridge.
When
more than
one
measurement
per
sensor
is
necessary
(lnstructions
7
and
9), excitation is
applied
separately
for
each
measurement.
(For
example, in Instruction
9
(6
wire full bridge),lhe
differential measurement
of
the
voltage drop
across the sensor
is
made with the
excitation
at
both
polarities;
excitation
is
again
applied
and
reversed
for the
measurement
of the
outpul
voltage.)
Instruction
8 applies an excitation
voltage,
delays a specified time, and
makes a differential
voltage
measurement, lf a
delay
of 0
is
specified,
the
inputs for
the differential
measurement are not
switched
for a second
integration
as is
normally
the case
(Section
13.2). The result stored is
the
voltage
measured.
Instruction
8
does not
have as
good
resolution
or common mode rejection
as the
ratiometric
bridge
measurement instructions" lt
does
provide
a very
rapid
means of making
bridge
measurements as
well as supplying
excitation
to circuitry
requiring differential
measurements. This
instruction
does
not
reverse
excitation.
A
1 before the excitation
channel number
(1X)
causes
the channel to be
incremented
with each repetition.
The
output of
lnstruction
8 is simply
the
voltage
measurement. When
8
is
used
to measure
a
full
bridge
(same
connections as
Instruction
6 in
Figure
13.5-1), the
result is Vl which equals
V,
(Rs/(Rs+Ra)
R2l(R1+R2)).
(ln
other
words, to
make
the output
the same
as
Instruction 6, use a
factor
of 10004/,
in the
multiplier.)
13-15
Commentaires sur ces manuels