Section 4. Troubleshooting
4. The OBS-3A and PC are not set to the same baud rate or communication
protocol (e.g. RS-232, USB, RS-485).
x Click
and check port settings on the serial port tab. The default
baud rate is 115.2 kb. If the PC is not set to this speed, follow the
steps in Section 3.4 to set it.
x If the OBS-3A still fails to respond, try changing PC speeds and
clicking
until communication is established (e.g. 57.6, 38.4, 19.6,
9.6 kb, etc.). If this fails, switch the PC back to 115.2 kb and go to the
next step.
x Reconnect the cable and try
.
x Replace the main batteries; see Section 2.2 and try
.
x If you have a survey cable, connect instrument to external power and
try
.
x Remove the unit from the pressure housing and press and release the
RESET button. Try
.
Power failed due to battery clip corrosion or a broken power wire.
Check for a broken red wire connecting the battery tube and circuit board.
Green powder or tarnish on the battery contact parts indicates salt-water
corrosion. Remove the electronics from the pressure housing. Pull battery-
clip-retainer pin out with needle-nose pliers and slide the clip from its track.
Clean the corroded surfaces of clip and track with a Scotch-brite
®
pad and
reassemble unit.
OBS or other sensor malfunction.
x Inspect for physical damage such as a broken or bent thermistor, a dirty
conductivity sensor, or an OBS sensor fouled with marine growth.
x Open unit and inspect for broken sensor and communication wires and
loose connectors (FIGURE 3.19-1).
x Check sensor power by starting Survey mode
and selecting all
sensors. Green LEDs should illuminate for installed sensor.
x If the depth sensor reads high and does not change, it may need to be
cleaned (see pressure-sensor maintenance, Section 5.2).
x If the sensors appear to be in working order, the digitizer or
microcontroller may be damaged. Such problems usually require factory
service.
4-2