Section 6. Calibration
Preparation
1. Clean containers and glassware with detergent and rinse with filtered
water.
2. Perform the calibration under fluorescent lighting.
3. Based on the material, select the appropriate sample duration from TABLE
6.2-1.
4. Open the calibration box with
and enter sediment concentration
values.
5. After each addition of sediment compute mg/l or ppm with the equations
given below.
6. Start the OBS-3A Utility and wake the OBS-3A. Click the mg/l or ppm
button.
TABLE 6.2-1. Sample Durations for
Sediment Calibrations
Sediment Seconds
Clay 10
Silt 20
Fine Sand 40
Sediment concentrations are calculated with the following equations:
mg/l
M
V
M
s
s
i
s
»
¼
º
«
¬
ª
;
ppm
M
M
i
s
M
s
Where:
M
s
= Mass (mg) of sediment in suspension
M
i
= Initial water mass, )(1 i kgVu
V
i
= Initial volume (L)
U
s
= Sediment density (usually ) l/mg1065.2
3
u
7. For the zero calibration point you will need a clean black 20” x 14” x 16”
container filled with clean tap water. A Rubbermaid plastic storage box is
suitable.
8. Add 2 l of filtered water to the 1 gallon container submerging the sensor at
least 5 cm; tap bubbles off container walls.
6-6