Section 11. Programming Resource Library
will respond, however, all other probes will remain active until the timeout
period expires.
Example:
Probe: Water Content
Power Usage:
Quiescent: 0.25 mA
Measurement: 120 mA
Measurement Time: 15 s
Active: 66 mA
Timeout: 15 s
Probes 1, 2, 3, and 4 are connected to SDI-12 / Control Port 1.
The time line in TABLE. Example Power Usage Profile for a Network of SDI-12
Probes (p. 119) shows a 35 second power usage profile example.
Table 15. Example Power Usage Profile for a Network of SDI-12 Probes
Sec Command
All
Probes
Awake
Time
Out
Expires
1 mA 2 mA 3mA 4mA
Total
mA
1 1M! Yes
120 66 66 66 318
2
120 66 66 66 318
14
120 66 66 66 318
15
Yes 120 66 66 66 318
16 1D0! Yes
66 66 66 66 264
17
66 66 66 66 264
29
66 66 66 66 264
30
Yes 66 66 66 66 264
31
0.25 0.25 0.25 0.25 1
35
0.25 0.25 0.25 0.25 1
119