SP Lite2 Silicon Pyranometer
Wiring for Examples 3 and 4
Colour Function CR1000 Channels CR10X Channels
Red Radiation Signal S.E. Channel 1 S.E. Channel 1
Blue Signal Reference AG
Black Shield G G
EXAMPLE 3. Sample Instructions used to Measure a Daily Total Flux
Density with a CR1000
'{CR1000}
'\\\\\\\\\\\\\\\\\\\\\\\\\ DECLARATIONS /////////////////////////
Public MJ_m2 As Long
'\\\\\\\\\\\\\\\\\\\\\\\\ OUTPUT SECTION ////////////////////////
DataTable(Total_Flux_Density,true,-1)
OpenInterval
DataInterval(0,1440,Min,10)
Totalize(1, MJ_m2, FP2, 0)
EndTable
'\\\\\\\\\\\\\\\\\\\\\\\\\\\ PROGRAM ////////////////////////////
BeginProg
Scan(5,Sec, 3, 0)
VoltSe(MJ_m2, 1, mV250, 1, False, 0, _60Hz, 0.00012, 0)
' Set negative values to zero
If (MJ_m2 < 0) Then
MJ_m2 = 0
EndIf
CallTable Total_Flux_Density
NextScan
EndProg
EXAMPLE 4. Sample Instructions used to Measure a Daily Total Flux
Density with a CR10X
;{CR10X}
;
*Table 1 Program
01: 5 Execution Interval (seconds)
1: Volt (SE) (P1)
1: 1 Reps
2: 24 250 mV 60 Hz Rejection Range ; **
3: 1 SE Channel ; *
4: 1 Loc [ MJ_m2 ] ; *
5: .00012 Multiplier ; ***
6: 0 Offset