QUALITY ASSURANCE/QUALITY CONTROL DOCUMENTATION SERIES
TITLE
SERVICING AND CALIBRATION OF OPTICAL MONITORING
DATALOGGERS
TYPE
STANDARD OPERATING PROCEDURE
NUMBER
4250
DATE
MARCH 1994
AUTHORIZATIONS
TITLE NAME SIGNATURE
ORIGINATOR James H. Wagner
PROJECT MANAGER James H. Wagner
PROGRAM MANAGER David L. Dietrich
QA MANAGER Gloria S. Mercer
OTHER
REVISION HISTORY
REVISION
NO.
CHANGE
DESCRIPTION
DATE AUTHORIZATIONS
1.0 Add responsibilities and equipment. May 1996
Vue de la page 0
1 2 3 4 5 6 ... 55 56

Résumé du contenu

Page 1

QUALITY ASSURANCE/QUALITY CONTROL DOCUMENTATION SERIESTITLESERVICING AND CALIBRATION OF OPTICAL MONITORINGDATALOGGERSTYPESTANDARD OPERATING PROCEDUREN

Page 2 - Page i of i

Number 4250-2000Revision 1.0Date MAY 1996Page 1 of 71.0 PURPOSE AND APPLICABILITYThis technical instruction (TI) describes procedures for servicing an

Page 3 - Page 1 of 5

Number 4250-2000Revision 1.0Date MAY 1996Page 2 of 7• Ensure that all datalogger servicing is documented and archived in accordance with theprocedures

Page 4 - Page 2 of 5

Number 4250-2000Revision 1.0Date MAY 1996Page 3 of 7• Replacement components as required• Medium screwdrivers (flat-blade and Phillips-head)• Battery

Page 5 - Page 3 of 5

Number 4250-2000Revision 1.0Date MAY 1996Page 4 of 7CAMPBELL 21X DATALOGGERSERVICING DOCUMENTATION FORMDate: ___________________________Datalogger S/N

Page 6 - Page 4 of 5

Number 4250-2000Revision 1.0Date MAY 1996Page 5 of 7After a few seconds delay, the memory check results will bedisplayed. If all memory is installed

Page 7 - Page 5 of 5

Number 4250-2000Revision 1.0Date MAY 1996Page 6 of 7Documentation Form. The correct reading for each port is shown,along with the manufacturers'

Page 8

Number 4250-2000Revision 1.0Date MAY 1996Page 7 of 7RECHARGE If the battery voltage is less than 11.76 volts, connect theBATTERY datalogger to the bat

Page 9 - Page i of i

QUALITY ASSURANCE/QUALITY CONTROL DOCUMENTATION SERIESTITLESERVICING AND CALIBRATION OF THE HANDAR 540A/570A DCPTYPETECHNICAL INSTRUCTIONNUMBER4250-20

Page 10 - Page 1 of 7

Number 4250-2010Revision 0.1Date MAY 1996Page i of iiTABLE OF CONTENTSSection Page1.0 PURPOSE AND APPLICABILITY 12.0 RESPONSIBILITIES 12.1 Project Man

Page 11 - Page 2 of 7

Number 4250-2010Revision 0.1Date MAY 1996Page ii of iiLIST OF FIGURES (CONTINUED)Figure Page4-2 Front Panel Configurations - Handar 540A/570A DCP 54-

Page 12 - Page 3 of 7

Number 4250 Revision 1.0 Date MAY 1996 Page i of iTABLE OF CONTENTSSection Page1.0 PURPOSE AND APPLICABILITY 12.0 RESPONS

Page 13 - SERVICING DOCUMENTATION FORM

Number 4250-2010Revision 0.1Date MAY 1996Page 1 of 371.0 PURPOSE AND APPLICABILITYThis technical instruction (TI) describes procedures for servicing a

Page 14 - Page 5 of 7

Number 4250-2010Revision 0.1Date MAY 1996Page 2 of 37• Coordinate with Handar for return, servicing, and recalibration of 540A/570A DCPsthat fail to o

Page 15 - Page 6 of 7

Number 4250-2010Revision 0.1Date MAY 1996Page 3 of 37• Electronic contacts cleaning fluid• Medium screwdrivers (flat-blade and Phillips-head)4.0 METHO

Page 16 - Page 7 of 7

Number 4250-2010Revision 0.1Date MAY 1996Page 4 of 37POST-FIELD INSPECTION CHECKLISTHANDAR 540A/570A DCPDCP S/N: ______________________________Site:

Page 17

Number 4250-2010Revision 0.1Date MAY 1996Page 5 of 37Figure 4-2. Front Panel Configurations - Handar 540A/570A DCP.

Page 18 - Page i of ii

Number 4250-2010Revision 0.1Date MAY 1996Page 6 of 374.1.2 Physical Inspection - ExternalINSPECT Inspect the outside of the DCP case thoroughly. Look

Page 19 - Page ii of ii

Number 4250-2010Revision 0.1Date MAY 1996Page 7 of 37Figure 4-3. DCP Component Diagram.

Page 20 - Page 1 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 8 of 37SWITCH When the "TERM" setup screen (see Figure 4-4) is displayed onBAUD RATE the compu

Page 21 - Page 2 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 9 of 37Figure 4-4. TERM Setup Screen.

Page 22 - Page 3 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 10 of 37Figure 4-5. "TERM" DCP Programming Screen (Program in DCP Memory).

Page 23 - HANDAR 540A/570A DCP

Number 4250 Revision 1.0 Date MAY 1996 Page 1 of 51.0 PURPOSE AND APPLICABILITYThis standard operating procedure (SOP) ou

Page 24 - Page 5 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 11 of 37Figure 4-6. TERM DCP Programming Screen (No Program in DCP Memory).

Page 25 - Page 6 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 12 of 37SWITCH Switch the DCP from "RUN" mode to "PROGRAM" mode byMODES entering ?.T

Page 26 - Page 7 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 13 of 37COMPARE The DCP data channel 04 reading (Rotronics AT output signal)READINGS must be adjusted by

Page 27 - Page 8 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 14 of 37ROUTINE SERVICING CHECKLISTHANDAR 540A/570A DCPDCP S/N: __________________________Date: ________

Page 28 - Page 9 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 15 of 37VERIFY Verify that the GOES primary channel select switches are set toGOES “900” and the seconda

Page 29 - Page 10 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 16 of 37CLEAN INTERIOR Clean the inside of the DCP with compressed air.OF DCPMODIFY 570A If this is a Ha

Page 30 - Page 11 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 17 of 37Figure 4-8. Handar 570A ADC Board - Component Locations.

Page 31 - Page 12 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 18 of 37Figure 4-9. Handar 540A Met Board - Component Locations.

Page 32 - Page 13 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 19 of 374.3 DCP PROGRAMMINGThe Handar 540A/570A DCP operational configuration is established through a u

Page 33 - ROUTINE SERVICING CHECKLIST

Number 4250-2010Revision 0.1Date MAY 1996Page 20 of 37Table 4-1DCP ID AssignmentsIMPROVE Transmissometer NetworkDCP-ID CHAN LOCATION TIME * RATE **FA4

Page 34 - Page 15 of 37

Number 4250 Revision 1.0 Date MAY 1996 Page 2 of 52.2 INSTRUMENT TECHNICIANThe instrument technician shall:• Perform all

Page 35 - Page 16 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 21 of 37EXECUTE Execute the Handar DCP communications program "TERM" fromPROGRAM the PC comput

Page 36 - Page 17 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 22 of 37• $ Forced Scan• # Forced Transmit (GOES Radio)Enter I to edit the station ID. The download pro

Page 37 - Page 18 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 23 of 37Enter M to select DCP channel 01. The display message will be "PCHANNEL 01."Enter N to

Page 38 - Page 19 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 24 of 37LABORATORY PERFORMANCE TESTINGHANDAR 540A/570A DCPDCP S/N: __________________________Date: _____

Page 39 - Page 20 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 25 of 374.4.3 Run Mode Timing ChecksVERIFY Verify that the GOES primary channel select switches are set

Page 40 - Page 21 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 26 of 37Set the power element of the wattmeter for the forward direction.Monitor the wattmeter reading a

Page 41 - Page 22 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 27 of 37APPENDIX AHANDAR 540A DCP CONFIGURATION PROGRAM - 540ROT.DCP

Page 42 - Page 23 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 28 of 37P ID FA43D0BEP STATION TIME 15:47:38P YEAR (XX) 94P DCP JULIAN DATE 055P GOES PRI XMT

Page 43

Number 4250-2010Revision 0.1Date MAY 1996Page 29 of 37P CHANNEL NO. 03P03 SENSOR TYPE 10P03 CARD SLOT # 06

Page 44 - Page 25 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 30 of 37P CHANNEL NO. 06P06 SENSOR TYPE 10P06 CARD SLOT # 06

Page 45 - Page 26 of 37

Number 4250 Revision 1.0 Date MAY 1996 Page 3 of 5• Reference thermometer (°C)• Replacement components as required• Batte

Page 46 - APPENDIX A

Number 4250-2010Revision 0.1Date MAY 1996Page 31 of 37P08 LOW DIFF LIMIT NO LIMITP CHANNEL NO. 09P09 SENSOR TYPE 10

Page 47 - Page 28 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 32 of 37APPENDIX BHANDAR 570A DCP CONFIGURATION PROGRAM - 570TROT.DCP

Page 48 - Page 29 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 33 of 37P ID FA43D0BEP STATION TIME 22:46:25P YEAR (XX) 94

Page 49 - Page 30 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 34 of 37P02 MEAS INTERVAL 01:00:00P02 START OF MEAS 17:30:00P02 LEVEL 1 MEAS TYPE 001

Page 50 - Page 31 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 35 of 37P05 SENSOR TYPE 10P05 SENSOR NAME TAG 10P05 CARD SLOT # 01

Page 51 - APPENDIX B

Number 4250-2010Revision 0.1Date MAY 1996Page 36 of 37P07 ZERO SCALE 000P07 MEAS INTERVAL 01:00:00P07 START OF MEAS 17:30:00

Page 52 - Page 33 of 37

Number 4250-2010Revision 0.1Date MAY 1996Page 37 of 37P CHANNEL NO. 10P10 SENSOR TYPE 12P10 SENSOR NAME TAG 12

Page 53 - Page 34 of 37

Number 4250 Revision 1.0 Date MAY 1996 Page 4 of 5• Frequency counter• Standard electronics laboratory small tools• Solte

Page 54 - Page 35 of 37

Number 4250 Revision 1.0 Date MAY 1996 Page 5 of 5• Routine laboratory servicing• DCP programming• Pre-field performance

Page 55 - Page 36 of 37

QUALITY ASSURANCE/QUALITY CONTROL DOCUMENTATION SERIESTITLESERVICING AND CALIBRATION OF CAMPBELL 21X DATALOGGERSTYPETECHNICAL INSTRUCTIONNUMBER4250-20

Page 56 - Page 37 of 37

Number 4250-2000Revision 1.0Date MAY 1996Page i of iTABLE OF CONTENTSSection Page1.0 PURPOSE AND APPLICABILITY 12.0 RESPONSIBILITIES 12.1 Project Mana