Campbell 21X Spécifications Page 27

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 56
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 26
Number 4250-2010
Revision 0.1
Date MAY 1996
Page 8 of 37
SWITCH When the "TERM" setup screen (see Figure 4-4) is displayed on
BAUD RATE the computer screen, press F2 to switch to the correct baud rate (300
baud).
CONNECT Connect the serial port of the computer to the DCP programming
COMPUTER port (see Figure 4-2 for location) using the DCP programming
TO DCP cable. After the connection is complete, the "TERM" programming
screen (Figure 4-5) should be displayed on the computer screen.
NOTE If the program is still in memory, the computer display should be
DISPLAY as shown in Figure 4-5.
If the computer display appears as in Figure 4-6, the program
has been lost from memory and the DCP performance and
timing checks should be terminated. Exit the "TERM"
program and turn the DCP off before initiating the servicing
procedures. Be sure and document that the program was no
longer in memory.
SYNCHRONIZE Verify that your watch is synchronized with WWV by calling the
TIME NIST WWV time transmission telephone (303/499-7111).
ENTER Obtain DCP ID and timing information by entering into the DCP
PARAMETERS the boldface character that precedes each of the following
parameters:
I Station ID
J DCP time
S Time remaining before next scan
T Time remaining before next transmission
RECORD Record each of the values from the ID and timing checks and the
VALUES correct (WWV) time when the DCP time check, [J], was performed.
COMPARE Compare the DCP times to the correct time to determine the timing
TIMES drift relative to previous measurements (at installation or from
recent transmissions - Refer to TI 4300-4000, Data Collection via
DCP (IMPROVE Protocol).
4.1.5 DCP A/D Converter Checks
CONNECT Connect a calibrated voltage source to the input of data channels
VOLTAGE 01, 02, and 03 of the DCP.
SOURCE
CONNECT Connect a calibrated Rotronic AT/RH sensor to the DCP's AT/RH
AT/RH SENSOR input connector.
Vue de la page 26
1 2 ... 22 23 24 25 26 27 28 29 30 31 32 ... 55 56

Commentaires sur ces manuels