Appendix B. SDI-12 Commands/Changing Settings
B-6
Table B-6. Example for Setting Power Operation Mode
Initial Command Response
0XSPOM+2!
Where,
0—sensor’s address;
2—the new power mode setting
(2 = auto).
00011<cr><If>
Where (from left to right),
0—sensor’s address;
001—the amount of time (in seconds)
that you must wait before sending
another data command;
1—the number of values that will be
placed in the buffer.
Subsequent Command Response
aXGPOM!
Where zero is the sensor address.
This is the get power mode
command.
0+2<cr><If>
Where,
0—sensor’s address;
2—the new power mode setting
(2 = auto).