Campbell TDR100 Manuel d'instructions Page 17

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 56
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 16
TDR100
11
(
)
L
L
w
v
=
⋅+
+
θ
max
.
.
0176
0114
2
with L the actual probe rod length, and, θ
v-max
the maximum expected
volumetric water content. Two m is added for the .5 m before the probe and
some distance after the probe end. For example, using a CS610 with 0.3 m
probe rod length in a soil with a porosity of 0.6 gives an estimated apparent
probe length of 4.04 m. Setting the Waveform Length to 4 m is recommended.
5. System Components: Datalogger Control
5.1 General
CR10X,
CR23X,
or
CR1000
FIGURE 5-1. TDR System Components
5.2 Datalogger
Campbell Scientific CR10X and CR23X dataloggers use Instruction 119 and
various other instructions to control the TDR100 measurement sequence and
store the resulting data. PC208W (version 3.1 or higher) or LoggerNet are
used to link the datalogger to a computer for data and program transfer.
Campbell Scientific CR1000 dataloggers use Instruction “TDR100” to control
the TDR100 measurement sequence and store the resulting data. PC400 or