TDR100
2
2. System Specifications
2.1 General
See the CR10X, CR23X, or CR1000 datalogger manuals for datalogger
specifications.
2.2 Power Consumption
2.2.1 TDR100
The current demand for the TDR100 during measurement is 270
milliamps.
When the TDR100 is controlled by a datalogger, a 35 second timer puts
the device in sleep mode requiring about 20 milliamps. After 35 seconds
in sleep mode, a timer puts the TDR100 in standby mode requiring about
2 milliamps.
When the TDR100 is controlled by PCTDR, a 60 second time-out puts
the device in low power mode requiring about 60 milliamps.
2.2.2 SDMX50
The quiescent current demand for the SDMX50 multiplexer is less than 1
milliamp.
Current demand during switching is approximately 90 milliamps.
All multiplexers of the same level switch simultaneously (see Figure 5-1).
Switching takes less than 1 second.
2.3 TDR100 Performance Specifications
pulse generator output 250 mV into 50 ohms
output impedance
50 ohms ± 1%
time response of combined pulse
generator and sampling circuit
300 picoseconds
pulse generator aberrations
±5% within first 10 nanoseconds
±0.5% after 10 nanoseconds
pulse length 14 microseconds
timing resolution 12.2 picoseconds
waveform sampling
range
resolution
20 to 2048 waveform values over
chosen length
distance
(V
p
= 1) time (1 way travel)
-2 to 2100 meters 0 to 7 microseconds
1.8 millimeter 6.1 picoseconds