TDR100
12
LoggerNet (version 3.0 or higher) are used to create and send the CRBasic
Program to the CR1000 datalogger.
5.3 TDR100
The TDR100 contains the pulse generator for the signal applied to a TDR
probe. The TDR100 also digitizes the reflection and applies numerical
algorithms for measuring volumetric water content or electrical conductivity.
The TDR100 communicates with the datalogger using SDM protocol or with a
computer using PCTDR and serial communications.
5.4 Power Supply
The system operates on 12 V power. A user supplied deep cycle 12 V lead
acid battery is commonly used in remote installations. Two terminal strip
adapters for the battery posts are provided with the ENCTDR100 (see Figure
5-2). These terminal strips will mount to wing nut battery posts found on most
deep cycle lead acid batteries. Installations with AC power available should
use it to continuously charge the system battery. Remote installation without
AC power should keep the battery charged with an SP10R or SP20R solar
panel. See the applications note at
ftp://ftp.campbellsci.com/pub/outgoing/apnotes/pow-sup.pdf for discussion of
power supplies.
FIGURE 5-2. Terminal Strip Adapters for Connections to Battery
Campbell Scientific recommends using datalogger switched 12 volts to power
the TDR100. This will provide power savings and will automatically reset the
TDR100 and provide automatic recovery from system malfunctions. This
practice can reduce loss of measurement data when a problem exists.
Typically the switched 12 volts is turned on at the beginning of the datalogger
program table that contains the TDR measurement instructions, and it is turned
off at the end of the table. CR10X and CR23X datalogger Instruction 86 can
be used to set a control port for the switched 12 volt power. PC208W or
LoggerNet HELP provide guidance for controlling switched 12 volts with both
the CR10X and CR23X. See datalogger programming example 1 in program
example section of this manual. CRBasic Instruction “SW12” can be used to
switch 12 volt power on the CR1000 datalogger. See programming example 6
in the program example section of this manual.